Preloaded Coated Filters
Cat. No. 225-9010
SKC coated lters are shipped preloaded in the casse es with end plugs and
shrink bands in place. No assembly is required; they are ready to use.
Method & Chemical: OSHA 102 for acetic anhydride
Filter & Coating
: 2 glass ber lters coated with veratrylamine and di-n-octyl
phthalate. Four-piece 37-mm casse e. Open-face sampling con guration.
Prior to Sampling: Store at 39.2 F (4 C) for up to 3 months.
Sample Stability: Store samples at 39.2 F (4 C) or 71.6 F (22 C) for up to
15 days.
† Custom order due to very limited shelf-life.
For information on other available
coated lters, request SKC Publication 1219.
Operating Instructions
863 Valley View Road, Eighty Four, PA 15330
Tel: 724-941-9701 Fax: 724-941-1369
www.skcinc.com
How to Use the SKC Preloaded Coated Filters
1. Select one coated lter casse e for calibrating the ow rate. A red plug
secures the inlet, which is clearly marked “inlet,” and a blue plug secures the
outlet. Set up the sampling train for calibration with the representative lter
casse e in line. For “closed-face” sampling, remove the plugs and connect the
casse e to the sampling train. For “open-face” sampling, remove the outlet
plug and the casse e inlet section then connect the casse e to the sampling
train. For details on se ing up a sampling train, refer to the SKC Sample Setup
Guide “Sampling Train - Filters and Cyclones” (SKC Publication 1166).
2. Calibrate the pump to the desired ow rate using a lm owmeter or
calibrated precision rotameter. For details on calibrating a pump, refer to the
SKC Sample Setup Guide “Calibrating a Pump Using a Film Flowmeter” (SKC
Publication 1163).
3. Replace the casse e used for calibration with a fresh coated lter casse e
for sample collection. A ach the casse e to a workers collar, as close to the
breathing zone as possible, by using a lter casse e holder (SKC Cat. No.
225-1). Sample for the speci ed time interval and record the time.
4. Remove the lter casse e at the end of the speci ed sampling period and
replace both end plugs (and the inlet if necessary). Recheck the ow rate using
the same casse e and owmeter used for calibration (in Step 2) to ensure that
the ow rate has not changed by more than 5%.
5. See Sample Stability on reverse side for sample storage. Appropriately package
filter cassettes and ship the samples, blanks, and all pertinent data to a
laboratory for analysis. Expedite shipping of samples to laboratory. Friday
shipments are not recommended.
Form 40039 Rev 1303
SKC Limited Warranty and Return Policy
SKC products are subject to the SKC Limited Warranty and Return Policy, which
provides SKC's sole liability and the buyer's exclusive remedy. To view the complete
SKC Limited Warranty and Return Policy, go to http://www.skcinc.com/warranty.asp.
Page view 0
1

Summary of Contents

Page 1 - Operating Instructions

Preloaded Coated FiltersCat. No. 225-9010SKC coated fi lters are shipped preloaded in the casse es with end plugs and shrink bands in place. No assem

Comments to this Manuals

No comments