1SKC Re-usable Parallel Particle Impactors (PPI)The patented† impaction-based SKC Re-usable Parallel Particle Impactor (PPI ) Samplers are designed to
2Principle of OperationSKC Re-usable PPI Samplers are impaction-based fi lter samplers that perform precise size-selection for either thoracic or resp
34. Using forceps, insert a 37-mm support pad and a 37-mm collection fi lter into the base plate.Inserting Impaction Substrates into the PPI1. Ensure
4Calibration and SamplingAs the particle load on the fi lter increases during sampling, the pressure drop will also increase. Therefore, use a compensa
53. Calibrate the sample pump to the specifi ed fl ow rate. See sample pump and calibrator operating instructions.4. Disconnect tubing and remove cali
6Removing the Collection Filter and Impaction Substrates1. Unscrew and remove the two screws that hold the inlet plate to the base plate.2. Li off
7CleaningFor optimum performance, the PPI inlet, impaction, and base plates should be cleaned a er fi ve runs or upon a noticeable buildup of material
8Ordering InformationPPI Samplers, require fi lter, substrates, and support Cat. No.Respirable PPI (gold), 2 L/min 225-380Thoracic PPI (blue), 2 L/m