Turn the ow adjust screw (Figure 3, #2) for the appropriate port
of the low ow holder until the desired ow rate is achieved.
Turn clockwise to decrease the ow.
Do not adjust the ow on the pump. Adjust the ow only by
using the ow adjust screw on the low ow holder.
Do not exceed 500 ml/min ow rate per tube.
When the desired ow is set on the
initial tube, place the pump in Hold
by pressing Flow and Battery Check.
Remove the calibrator from the tube
and connect to the exposed end of
the next sorbent tube. Press Flow and
Battery Check and repeat the ow
adjustment process until all tubes are
ow calibrated. Changing the ow on
one tube will not affect the ow rate through the remaining tubes.
Do not exceed 500 ml/min ow rate per tube.
For tri and quad models, rst rotate each anti-tamper cover
(Figures 3 and 4) to expose the ow adjust screws, then
adjust the appropriate screw until the calibrator indicates the
desired ow.
When the ow rate is set for each tube, press Flow and Battery
Check to place the pump in Hold and disconnect the calibrator.
Replace the sampling media used for calibration with unexposed
media for sample collection. Use a protective tube cover to
prevent tube breakage.
If sampling with fewer tubes than number of ports, insert unopened
sorbent tubes in the empty ports to seal them.
2
(cont’d)
Screw
Screw
Figure 4
Cut-away of Tri/Quad Low
Flow Holder